VOLUME 1: MARS EXPRESS ASPERA3 EDR ELS DATA View Errata Order Volume Toggle viewing into brief mode
MEX-M-ASPERA3-2-EDR-ELS-EXT3-V1.0 PDS 3      CERTIFIED
Start Time: 2010-01-02 07:50:15 - Stop Time: 2013-01-01 13:29:19
Extension 3 raw data from the Electron Spectrometer (ELS) instrument of the Analyzer of Space Plasmas and Energetic Atoms experiment, 3rd edition (ASPERA-3), aboard the Mars Express spacecraft. The ASPERA-3 Principal Investigator is Dr. Rickard Lundin, Swedish Institute of Space Physics (IRF), Kiruna, Sweden. The primary source for ASPERA-3 data     more ...
Extension 3 raw data from the Electron Spectrometer (ELS) instrument of the Analyzer of Space Plasmas and Energetic Atoms experiment, 3rd edition (ASPERA-3), aboard the Mars Express spacecraft. The ASPERA-3 Principal Investigator is Dr. Rickard Lundin, Swedish Institute of Space Physics (IRF), Kiruna, Sweden. The primary source for ASPERA-3 data is the ESA Planetary Science Archive (PSA).    less ...
Citiation Information
Lundin, R., S. Barabash, D. Winningham, R. Frahm, MARS EXPRESS ASPERA-3 RAW EDR ELECTRON SPECTROMTR EXT3 V1.0, MEX-M-ASPERA3-2-EDR-ELS-EXT3-V1.0, ESA Planetary Science Archive (PSA), NASA Planetary Data System (PDS), 2005
Top/ DATA/ 2011/ 152_181JUN / click here to download folders and files
ELSSCIH20111520415C_ACCS01
Start Time: 2011-152T04:15:23.754 Stop Time: 2011-152T09:21:16.816
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111521115C_ACCS01
Start Time: 2011-152T11:15:12.443 Stop Time: 2011-152T16:20:33.093
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111521814C_ACCS01
Start Time: 2011-152T18:14:28.720 Stop Time: 2011-152T21:05:57.023
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111522110C_ACCS01
Start Time: 2011-152T21:10:09.024 Stop Time: 2011-152T23:20:16.218
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111530329C_ACCS01
Start Time: 2011-153T03:29:39.567 Stop Time: 2011-153T06:19:57.271
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111530813C_ACCS01
Start Time: 2011-153T08:13:20.837 Stop Time: 2011-153T13:19:13.741
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111531512C_ACCS01
Start Time: 2011-153T15:12:37.273 Stop Time: 2011-153T20:19:01.859
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111532211C_ACCS01
Start Time: 2011-153T22:11:53.359 Stop Time: 2011-154T03:18:49.428
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111540511C_ACCS01
Start Time: 2011-154T05:11:08.868 Stop Time: 2011-154T10:18:37.301
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111541427C_ACCS01
Start Time: 2011-154T14:27:50.582 Stop Time: 2011-154T17:18:08.102
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111541910C_ACCS01
Start Time: 2011-154T19:10:27.575 Stop Time: 2011-155T00:17:56.858
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111550209C_ACCS01
Start Time: 2011-155T02:09:44.266 Stop Time: 2011-155T05:02:55.312
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111550507C_ACCS01
Start Time: 2011-155T05:07:07.751 Stop Time: 2011-155T07:17:14.726
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111551126C_ACCS01
Start Time: 2011-155T11:26:31.444 Stop Time: 2011-155T14:16:49.807
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111551608C_ACCS01
Start Time: 2011-155T16:08:37.184 Stop Time: 2011-155T21:16:38.059
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111552307C_ACCS01
Start Time: 2011-155T23:07:53.432 Stop Time: 2011-156T02:01:48.360
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111560206C_ACCS01
Start Time: 2011-156T02:06:00.424 Stop Time: 2011-156T04:16:06.772
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111560825C_ACCS01
Start Time: 2011-156T08:25:37.431 Stop Time: 2011-156T11:15:55.699
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111561306C_ACCS01
Start Time: 2011-156T13:06:38.953 Stop Time: 2011-156T18:15:44.515
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111562006C_ACCS01
Start Time: 2011-156T20:06:27.765 Stop Time: 2011-157T01:15:01.702
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111570305C_ACCS01
Start Time: 2011-157T03:05:44.953 Stop Time: 2011-157T08:14:50.539
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111571005C_ACCS01
Start Time: 2011-157T10:05:01.728 Stop Time: 2011-157T12:59:51.246
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111571304C_ACCS01
Start Time: 2011-157T13:04:03.528 Stop Time: 2011-157T15:14:10.331
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111571923C_ACCS01
Start Time: 2011-157T19:23:42.575 Stop Time: 2011-157T22:14:00.375
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111580003C_ACCS01
Start Time: 2011-158T00:03:39.504 Stop Time: 2011-158T05:13:49.499
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111580703C_ACCS01
Start Time: 2011-158T07:03:28.657 Stop Time: 2011-158T12:13:07.272
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111581402C_ACCS01
Start Time: 2011-158T14:02:46.334 Stop Time: 2011-158T16:58:33.722
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111581702C_ACCS01
Start Time: 2011-158T17:02:46.098 Stop Time: 2011-158T19:12:52.980
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111582322C_ACCS01
Start Time: 2011-158T23:22:17.472 Stop Time: 2011-159T02:12:35.459
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111590401C_ACCS01
Start Time: 2011-159T04:01:42.524 Stop Time: 2011-159T09:11:53.365
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111591101C_ACCS01
Start Time: 2011-159T11:01:00.460 Stop Time: 2011-159T16:11:42.798
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111591800C_ACCS01
Start Time: 2011-159T18:00:17.766 Stop Time: 2011-159T23:11:32.842
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111600059C_ACCS01
Start Time: 2011-160T00:59:35.750 Stop Time: 2011-160T06:10:49.649
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111601020C_ACCS01
Start Time: 2011-160T10:20:32.453 Stop Time: 2011-160T13:10:50.880
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111601458C_ACCS01
Start Time: 2011-160T14:58:53.820 Stop Time: 2011-160T20:10:08.287
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111602158C_ACCS01
Start Time: 2011-160T21:58:11.250 Stop Time: 2011-161T03:09:57.121
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111610457C_ACCS01
Start Time: 2011-161T04:57:28.057 Stop Time: 2011-161T10:09:46.467
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111611418C_ACCS01
Start Time: 2011-161T14:18:53.459 Stop Time: 2011-161T17:09:11.668
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111611856C_ACCS01
Start Time: 2011-161T18:56:10.512 Stop Time: 2011-162T00:09:01.609
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111620156C_ACCS01
Start Time: 2011-162T01:56:00.484 Stop Time: 2011-162T07:08:19.259
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111620855C_ACCS01
Start Time: 2011-162T08:55:18.104 Stop Time: 2011-162T11:53:50.064
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111621158C_ACCS01
Start Time: 2011-162T11:58:02.033 Stop Time: 2011-162T14:08:08.759
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111621816C_ACCS01
Start Time: 2011-162T18:16:26.458 Stop Time: 2011-162T21:07:49.539
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111622254C_ACCS01
Start Time: 2011-162T22:54:16.321 Stop Time: 2011-163T04:07:07.009
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111630553C_ACCS01
Start Time: 2011-163T05:53:33.820 Stop Time: 2011-163T11:06:56.504
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111631252C_ACCS01
Start Time: 2011-163T12:52:51.192 Stop Time: 2011-163T15:52:11.123
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111631556C_ACCS01
Start Time: 2011-163T15:56:23.155 Stop Time: 2011-163T18:05:57.693
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111632215C_ACCS01
Start Time: 2011-163T22:15:59.465 Stop Time: 2011-164T01:06:17.142
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111640251C_ACCS01
Start Time: 2011-164T02:51:39.798 Stop Time: 2011-164T08:06:06.891
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111640950C_ACCS01
Start Time: 2011-164T09:50:57.451 Stop Time: 2011-164T15:05:24.354
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111641650C_ACCS01
Start Time: 2011-164T16:50:46.980 Stop Time: 2011-164T22:05:13.577
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111642350C_ACCS01
Start Time: 2011-164T23:50:04.108 Stop Time: 2011-165T02:50:23.194
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111650254C_ACCS01
Start Time: 2011-165T02:54:35.601 Stop Time: 2011-165T05:04:42.511
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111650914C_ACCS01
Start Time: 2011-165T09:14:05.469 Stop Time: 2011-165T12:04:23.923
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111651348C_ACCS01
Start Time: 2011-165T13:48:42.423 Stop Time: 2011-165T19:04:14.173
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111652048C_ACCS01
Start Time: 2011-165T20:48:00.610 Stop Time: 2011-165T23:49:20.159
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111652353C_ACCS01
Start Time: 2011-165T23:53:32.441 Stop Time: 2011-166T02:03:39.353
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111660613C_ACCS01
Start Time: 2011-166T06:13:02.589 Stop Time: 2011-166T09:03:20.329
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111661047C_ACCS01
Start Time: 2011-166T10:47:06.765 Stop Time: 2011-166T16:03:09.423
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111661746C_ACCS01
Start Time: 2011-166T17:46:23.732 Stop Time: 2011-166T23:02:27.471
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111670045C_ACCS01
Start Time: 2011-167T00:45:41.876 Stop Time: 2011-167T03:47:50.927
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111670352C_ACCS01
Start Time: 2011-167T03:52:03.147 Stop Time: 2011-167T06:02:10.825
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111671010C_ACCS01
Start Time: 2011-167T10:10:29.342 Stop Time: 2011-167T13:01:51.926
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111671444C_ACCS01
Start Time: 2011-167T14:44:34.204 Stop Time: 2011-167T20:01:42.300
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111672144C_ACCS01
Start Time: 2011-167T21:44:24.642 Stop Time: 2011-168T03:01:00.110
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111680443C_ACCS01
Start Time: 2011-168T04:43:42.422 Stop Time: 2011-168T10:00:49.544
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111681409C_ACCS01
Start Time: 2011-168T14:09:10.475 Stop Time: 2011-168T17:00:32.653
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111681842C_ACCS01
Start Time: 2011-168T18:42:42.933 Stop Time: 2011-168T23:59:51.495
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111690142C_ACCS01
Start Time: 2011-169T01:42:01.742 Stop Time: 2011-169T06:59:41.958
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111690841C_ACCS01
Start Time: 2011-169T08:41:20.147 Stop Time: 2011-169T13:59:32.579
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111691808C_ACCS01
Start Time: 2011-169T18:08:45.473 Stop Time: 2011-169T20:59:03.962
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111692240C_ACCS01
Start Time: 2011-169T22:40:10.086 Stop Time: 2011-170T03:58:53.928
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111700540C_ACCS01
Start Time: 2011-170T05:40:00.050 Stop Time: 2011-170T10:58:12.203
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111701507C_ACCS01
Start Time: 2011-170T15:07:47.605 Stop Time: 2011-170T17:58:05.312
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111701938C_ACCS01
Start Time: 2011-170T19:38:39.374 Stop Time: 2011-171T00:57:23.095
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111710237C_ACCS01
Start Time: 2011-171T02:37:57.185 Stop Time: 2011-171T05:42:56.672
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111710547C_ACCS01
Start Time: 2011-171T05:47:08.579 Stop Time: 2011-171T07:57:14.771
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111711206C_ACCS01
Start Time: 2011-171T12:06:38.595 Stop Time: 2011-171T14:56:56.337
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111711636C_ACCS01
Start Time: 2011-171T16:36:58.306 Stop Time: 2011-171T21:56:46.930
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111712336C_ACCS01
Start Time: 2011-171T23:36:16.866 Stop Time: 2011-172T04:56:05.112
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111720905C_ACCS01
Start Time: 2011-172T09:05:38.604 Stop Time: 2011-172T11:55:56.998
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111721335C_ACCS01
Start Time: 2011-172T13:35:26.967 Stop Time: 2011-172T18:55:47.468
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111722034C_ACCS01
Start Time: 2011-172T20:34:45.306 Stop Time: 2011-173T01:00:58.258
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111731350C_ACCS01
Start Time: 2011-173T13:50:34.561 Stop Time: 2011-173T15:54:15.189
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111732003C_ACCS01
Start Time: 2011-173T20:03:52.483 Stop Time: 2011-173T22:54:10.192
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111740032C_ACCS01
Start Time: 2011-174T00:32:36.003 Stop Time: 2011-174T05:54:00.661
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111740731C_ACCS01
Start Time: 2011-174T07:31:54.408 Stop Time: 2011-174T10:39:03.775
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111741043C_ACCS01
Start Time: 2011-174T10:43:15.464 Stop Time: 2011-174T12:53:22.282
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111741702C_ACCS01
Start Time: 2011-174T17:02:47.599 Stop Time: 2011-174T19:53:05.965
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111742130C_ACCS01
Start Time: 2011-174T21:30:59.713 Stop Time: 2011-175T02:52:24.737
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111750430C_ACCS01
Start Time: 2011-175T04:30:18.483 Stop Time: 2011-175T09:52:15.549
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111751400C_ACCS01
Start Time: 2011-175T14:00:34.469 Stop Time: 2011-175T16:51:56.643
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111751828C_ACCS01
Start Time: 2011-175T18:28:46.328 Stop Time: 2011-175T23:51:47.486
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111760128C_ACCS01
Start Time: 2011-176T01:28:37.138 Stop Time: 2011-176T06:51:06.095
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111760827C_ACCS01
Start Time: 2011-176T08:27:55.717 Stop Time: 2011-176T11:36:28.580
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111761140C_ACCS01
Start Time: 2011-176T11:40:41.143 Stop Time: 2011-176T13:50:48.555
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111761800C_ACCS01
Start Time: 2011-176T18:00:13.490 Stop Time: 2011-176T20:50:31.444
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111762226C_ACCS01
Start Time: 2011-176T22:26:49.098 Stop Time: 2011-177T03:49:50.786
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111770526C_ACCS01
Start Time: 2011-177T05:26:08.346 Stop Time: 2011-177T10:49:42.408
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111771225C_ACCS01
Start Time: 2011-177T12:25:27.908 Stop Time: 2011-177T17:49:01.619
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111772158C_ACCS01
Start Time: 2011-177T21:58:46.615 Stop Time: 2011-178T00:49:04.295
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111780224C_ACCS01
Start Time: 2011-178T02:24:17.731 Stop Time: 2011-178T05:34:20.540
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111780538C_ACCS01
Start Time: 2011-178T05:38:33.009 Stop Time: 2011-178T07:48:39.416
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111781157C_ACCS01
Start Time: 2011-178T11:57:59.477 Stop Time: 2011-178T14:54:43.314
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111781623C_ACCS01
Start Time: 2011-178T16:23:31.998 Stop Time: 2011-178T21:42:48.214
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111782323C_ACCS01
Start Time: 2011-178T23:23:22.274 Stop Time: 2011-179T04:42:37.743
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111790623C_ACCS01
Start Time: 2011-179T06:23:11.771 Stop Time: 2011-179T11:41:54.921
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111791323C_ACCS01
Start Time: 2011-179T13:23:01.046 Stop Time: 2011-179T18:41:44.889
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111792256C_ACCS01
Start Time: 2011-179T22:56:11.615 Stop Time: 2011-180T01:41:39.881
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111800323C_ACCS01
Start Time: 2011-180T03:23:18.070 Stop Time: 2011-180T08:40:57.912
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111801023C_ACCS01
Start Time: 2011-180T10:23:08.192 Stop Time: 2011-180T15:40:48.077
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111801722C_ACCS01
Start Time: 2011-180T17:22:58.294 Stop Time: 2011-180T22:39:33.833
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111810022C_ACCS01
Start Time: 2011-181T00:22:48.238 Stop Time: 2011-181T03:30:25.453
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111810334C_ACCS01
Start Time: 2011-181T03:34:37.735 Stop Time: 2011-181T05:39:55.019
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111810954C_ACCS01
Start Time: 2011-181T09:54:15.496 Stop Time: 2011-181T12:39:43.388
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111811422C_ACCS01
Start Time: 2011-181T14:22:57.765 Stop Time: 2011-181T19:39:00.607
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20111812122C_ACCS01
Start Time: 2011-181T21:22:47.045 Stop Time: 2011-182T02:38:50.539
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111520415C_ACCS01
Start Time: 2011-152T04:15:25.629 Stop Time: 2011-152T09:21:18.910
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111521115C_ACCS01
Start Time: 2011-152T11:15:14.318 Stop Time: 2011-152T16:20:35.187
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111521814C_ACCS01
Start Time: 2011-152T18:14:30.595 Stop Time: 2011-152T21:05:59.116
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111522109C_ACCS01
Start Time: 2011-152T21:09:38.648 Stop Time: 2011-152T23:20:18.311
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111530329C_ACCS01
Start Time: 2011-153T03:29:41.442 Stop Time: 2011-153T06:19:59.365
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111530813C_ACCS01
Start Time: 2011-153T08:13:22.712 Stop Time: 2011-153T13:19:15.834
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111531512C_ACCS01
Start Time: 2011-153T15:12:39.148 Stop Time: 2011-153T20:19:03.953
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111532211C_ACCS01
Start Time: 2011-153T22:11:55.234 Stop Time: 2011-154T03:18:51.521
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111540511C_ACCS01
Start Time: 2011-154T05:11:10.743 Stop Time: 2011-154T10:18:39.395
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111541427C_ACCS01
Start Time: 2011-154T14:27:52.457 Stop Time: 2011-154T17:18:10.195
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111541910C_ACCS01
Start Time: 2011-154T19:10:29.450 Stop Time: 2011-155T00:17:58.952
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111550209C_ACCS01
Start Time: 2011-155T02:09:46.141 Stop Time: 2011-155T05:02:57.406
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111550506C_ACCS01
Start Time: 2011-155T05:06:37.188 Stop Time: 2011-155T07:17:16.819
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111551126C_ACCS01
Start Time: 2011-155T11:26:33.319 Stop Time: 2011-155T14:16:51.901
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111551608C_ACCS01
Start Time: 2011-155T16:08:39.059 Stop Time: 2011-155T21:16:40.152
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111552307C_ACCS01
Start Time: 2011-155T23:07:55.307 Stop Time: 2011-156T02:01:50.454
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111560205C_ACCS01
Start Time: 2011-156T02:05:30.048 Stop Time: 2011-156T04:16:08.866
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111560825C_ACCS01
Start Time: 2011-156T08:25:39.306 Stop Time: 2011-156T11:15:57.792
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111561306C_ACCS01
Start Time: 2011-156T13:06:40.828 Stop Time: 2011-156T18:15:46.609
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111562006C_ACCS01
Start Time: 2011-156T20:06:29.640 Stop Time: 2011-157T01:15:03.795
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111570305C_ACCS01
Start Time: 2011-157T03:05:46.828 Stop Time: 2011-157T08:14:52.633
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111571005C_ACCS01
Start Time: 2011-157T10:05:03.603 Stop Time: 2011-157T12:59:53.340
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111571303C_ACCS01
Start Time: 2011-157T13:03:33.153 Stop Time: 2011-157T15:14:12.424
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111571923C_ACCS01
Start Time: 2011-157T19:23:44.450 Stop Time: 2011-157T22:14:02.469
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111580003C_ACCS01
Start Time: 2011-158T00:03:41.379 Stop Time: 2011-158T05:13:51.592
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111580703C_ACCS01
Start Time: 2011-158T07:03:30.532 Stop Time: 2011-158T12:13:09.366
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111581402C_ACCS01
Start Time: 2011-158T14:02:48.209 Stop Time: 2011-158T16:58:35.816
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111581702C_ACCS01
Start Time: 2011-158T17:02:15.723 Stop Time: 2011-158T19:12:55.073
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111582322C_ACCS01
Start Time: 2011-158T23:22:19.347 Stop Time: 2011-159T02:12:37.553
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111590401C_ACCS01
Start Time: 2011-159T04:01:44.399 Stop Time: 2011-159T09:11:55.459
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111591101C_ACCS01
Start Time: 2011-159T11:01:02.335 Stop Time: 2011-159T16:11:44.892
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111591800C_ACCS01
Start Time: 2011-159T18:00:19.641 Stop Time: 2011-159T23:11:34.936
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111600059C_ACCS01
Start Time: 2011-160T00:59:37.625 Stop Time: 2011-160T06:10:51.743
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111601020C_ACCS01
Start Time: 2011-160T10:20:34.328 Stop Time: 2011-160T13:10:52.974
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111601458C_ACCS01
Start Time: 2011-160T14:58:55.695 Stop Time: 2011-160T20:10:10.381
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111602158C_ACCS01
Start Time: 2011-160T21:58:13.125 Stop Time: 2011-161T03:09:59.215
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111610457C_ACCS01
Start Time: 2011-161T04:57:29.932 Stop Time: 2011-161T10:09:48.561
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111611418C_ACCS01
Start Time: 2011-161T14:18:55.334 Stop Time: 2011-161T17:09:13.762
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111611856C_ACCS01
Start Time: 2011-161T18:56:12.387 Stop Time: 2011-162T00:09:03.703
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111620156C_ACCS01
Start Time: 2011-162T01:56:02.359 Stop Time: 2011-162T07:08:21.353
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111620855C_ACCS01
Start Time: 2011-162T08:55:19.979 Stop Time: 2011-162T11:53:52.157
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111621157C_ACCS01
Start Time: 2011-162T11:57:31.658 Stop Time: 2011-162T14:08:10.853
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111621816C_ACCS01
Start Time: 2011-162T18:16:28.333 Stop Time: 2011-162T21:07:51.633
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111622254C_ACCS01
Start Time: 2011-162T22:54:18.196 Stop Time: 2011-163T04:07:09.103
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111630553C_ACCS01
Start Time: 2011-163T05:53:35.695 Stop Time: 2011-163T11:06:58.598
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111631252C_ACCS01
Start Time: 2011-163T12:52:53.067 Stop Time: 2011-163T15:52:13.216
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111631555C_ACCS01
Start Time: 2011-163T15:55:52.780 Stop Time: 2011-163T18:05:59.786
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111632216C_ACCS01
Start Time: 2011-163T22:16:01.340 Stop Time: 2011-164T01:06:19.236
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111640251C_ACCS01
Start Time: 2011-164T02:51:41.673 Stop Time: 2011-164T08:06:08.984
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111640950C_ACCS01
Start Time: 2011-164T09:50:59.326 Stop Time: 2011-164T15:05:26.448
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111641650C_ACCS01
Start Time: 2011-164T16:50:48.855 Stop Time: 2011-164T22:05:15.670
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111642350C_ACCS01
Start Time: 2011-164T23:50:05.983 Stop Time: 2011-165T02:50:25.288
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111650254C_ACCS01
Start Time: 2011-165T02:54:05.257 Stop Time: 2011-165T05:04:44.605
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111650914C_ACCS01
Start Time: 2011-165T09:14:07.344 Stop Time: 2011-165T12:04:26.016
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111651348C_ACCS01
Start Time: 2011-165T13:48:44.298 Stop Time: 2011-165T19:04:16.267
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111652048C_ACCS01
Start Time: 2011-165T20:48:02.485 Stop Time: 2011-165T23:49:22.253
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111652353C_ACCS01
Start Time: 2011-165T23:53:02.066 Stop Time: 2011-166T02:03:41.446
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111660613C_ACCS01
Start Time: 2011-166T06:13:04.464 Stop Time: 2011-166T09:03:22.423
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111661047C_ACCS01
Start Time: 2011-166T10:47:08.640 Stop Time: 2011-166T16:03:11.516
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111661746C_ACCS01
Start Time: 2011-166T17:46:25.607 Stop Time: 2011-166T23:02:29.565
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111670045C_ACCS01
Start Time: 2011-167T00:45:43.751 Stop Time: 2011-167T03:47:53.021
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111670351C_ACCS01
Start Time: 2011-167T03:51:32.771 Stop Time: 2011-167T06:02:12.918
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111671010C_ACCS01
Start Time: 2011-167T10:10:31.217 Stop Time: 2011-167T13:01:54.020
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111671444C_ACCS01
Start Time: 2011-167T14:44:36.079 Stop Time: 2011-167T20:01:44.393
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111672144C_ACCS01
Start Time: 2011-167T21:44:26.517 Stop Time: 2011-168T03:01:02.204
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111680443C_ACCS01
Start Time: 2011-168T04:43:44.297 Stop Time: 2011-168T10:00:51.638
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111681409C_ACCS01
Start Time: 2011-168T14:09:12.350 Stop Time: 2011-168T17:00:34.747
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111681842C_ACCS01
Start Time: 2011-168T18:42:44.808 Stop Time: 2011-168T23:59:53.589
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111690142C_ACCS01
Start Time: 2011-169T01:42:03.617 Stop Time: 2011-169T06:59:44.052
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111690841C_ACCS01
Start Time: 2011-169T08:41:22.022 Stop Time: 2011-169T13:59:34.673
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111691808C_ACCS01
Start Time: 2011-169T18:08:47.348 Stop Time: 2011-169T20:59:06.056
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111692240C_ACCS01
Start Time: 2011-169T22:40:11.961 Stop Time: 2011-170T03:58:56.022
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111700540C_ACCS01
Start Time: 2011-170T05:40:01.925 Stop Time: 2011-170T10:58:14.296
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111701507C_ACCS01
Start Time: 2011-170T15:07:49.480 Stop Time: 2011-170T17:58:07.406
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111701938C_ACCS01
Start Time: 2011-170T19:38:41.249 Stop Time: 2011-171T00:57:25.188
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111710237C_ACCS01
Start Time: 2011-171T02:37:59.060 Stop Time: 2011-171T05:42:58.766
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111710546C_ACCS01
Start Time: 2011-171T05:46:38.204 Stop Time: 2011-171T07:57:16.865
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111711206C_ACCS01
Start Time: 2011-171T12:06:40.470 Stop Time: 2011-171T14:56:58.430
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111711637C_ACCS01
Start Time: 2011-171T16:37:00.181 Stop Time: 2011-171T21:56:49.024
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111712336C_ACCS01
Start Time: 2011-171T23:36:18.741 Stop Time: 2011-172T04:56:07.206
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111720905C_ACCS01
Start Time: 2011-172T09:05:40.479 Stop Time: 2011-172T11:55:59.092
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111721335C_ACCS01
Start Time: 2011-172T13:35:28.842 Stop Time: 2011-172T18:55:49.562
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111722034C_ACCS01
Start Time: 2011-172T20:34:47.181 Stop Time: 2011-173T01:01:00.352
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111731350C_ACCS01
Start Time: 2011-173T13:50:36.436 Stop Time: 2011-173T15:54:17.283
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111732003C_ACCS01
Start Time: 2011-173T20:03:54.358 Stop Time: 2011-173T22:54:12.286
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111740032C_ACCS01
Start Time: 2011-174T00:32:37.878 Stop Time: 2011-174T05:54:02.754
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111740731C_ACCS01
Start Time: 2011-174T07:31:56.283 Stop Time: 2011-174T10:39:05.869
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111741042C_ACCS01
Start Time: 2011-174T10:42:45.088 Stop Time: 2011-174T12:53:24.376
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111741702C_ACCS01
Start Time: 2011-174T17:02:49.474 Stop Time: 2011-174T19:53:08.058
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111742131C_ACCS01
Start Time: 2011-174T21:31:01.588 Stop Time: 2011-175T02:52:26.831
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111750430C_ACCS01
Start Time: 2011-175T04:30:20.358 Stop Time: 2011-175T09:52:17.642
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111751400C_ACCS01
Start Time: 2011-175T14:00:36.344 Stop Time: 2011-175T16:51:58.737
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111751828C_ACCS01
Start Time: 2011-175T18:28:48.203 Stop Time: 2011-175T23:51:49.580
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111760128C_ACCS01
Start Time: 2011-176T01:28:39.013 Stop Time: 2011-176T06:51:08.189
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111760827C_ACCS01
Start Time: 2011-176T08:27:57.592 Stop Time: 2011-176T11:36:30.673
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111761140C_ACCS01
Start Time: 2011-176T11:40:10.768 Stop Time: 2011-176T13:50:50.649
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111761800C_ACCS01
Start Time: 2011-176T18:00:15.365 Stop Time: 2011-176T20:50:33.538
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111762226C_ACCS01
Start Time: 2011-176T22:26:50.973 Stop Time: 2011-177T03:49:52.880
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111770526C_ACCS01
Start Time: 2011-177T05:26:10.221 Stop Time: 2011-177T10:49:44.502
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111771225C_ACCS01
Start Time: 2011-177T12:25:29.783 Stop Time: 2011-177T17:49:03.713
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111772158C_ACCS01
Start Time: 2011-177T21:58:48.490 Stop Time: 2011-178T00:49:06.388
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111780224C_ACCS01
Start Time: 2011-178T02:24:19.606 Stop Time: 2011-178T05:34:22.634
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111780538C_ACCS01
Start Time: 2011-178T05:38:02.634 Stop Time: 2011-178T07:48:41.510
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111781158C_ACCS01
Start Time: 2011-178T11:58:01.352 Stop Time: 2011-178T14:54:45.407
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111781623C_ACCS01
Start Time: 2011-178T16:23:33.873 Stop Time: 2011-178T21:42:50.308
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111782323C_ACCS01
Start Time: 2011-178T23:23:24.149 Stop Time: 2011-179T04:42:39.836
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111790623C_ACCS01
Start Time: 2011-179T06:23:13.646 Stop Time: 2011-179T11:41:57.015
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111791323C_ACCS01
Start Time: 2011-179T13:23:02.921 Stop Time: 2011-179T18:41:46.982
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111792256C_ACCS01
Start Time: 2011-179T22:56:13.490 Stop Time: 2011-180T01:41:41.975
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111800323C_ACCS01
Start Time: 2011-180T03:23:19.945 Stop Time: 2011-180T08:41:00.006
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111801023C_ACCS01
Start Time: 2011-180T10:23:10.067 Stop Time: 2011-180T15:40:50.171
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111801723C_ACCS01
Start Time: 2011-180T17:23:00.169 Stop Time: 2011-180T22:39:35.927
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111810022C_ACCS01
Start Time: 2011-181T00:22:50.113 Stop Time: 2011-181T03:30:27.547
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111810334C_ACCS01
Start Time: 2011-181T03:34:07.391 Stop Time: 2011-181T05:39:57.113
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111810954C_ACCS01
Start Time: 2011-181T09:54:17.371 Stop Time: 2011-181T12:39:45.482
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111811422C_ACCS01
Start Time: 2011-181T14:22:59.640 Stop Time: 2011-181T19:39:02.700
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20111812122C_ACCS01
Start Time: 2011-181T21:22:48.920 Stop Time: 2011-182T02:38:52.633
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ICON LEGEND
For links to data holdings various icons are used to represent services or actions you can take. Those icons are:
Order the volume (CD-ROM/DVD) containing the listed data. The volume will be mailed to you. The volume (CD-ROM/DVD) contains some errors which are documented in an errata log.
Launch a custom web browser for the resource. Launch a map based browser for the resource.
Toggle viewing into brief mode (shorter descriptions are displayed). Toggle viewing into expanded mode (longer descriptions are displayed).
Download the volume containing the listed data.


NEW RELEASES

More.....

PDS RESOURCES

Archive Planning Guide
      Individuals
      Missions
Data Dictionary Search
      Lookup Tool
      Document
PDS4 Standards Documents
PDS Subscription Service

PDS3 Standards

OTHER RESOURCES

ADS Search
Science Data Licenses

Privacy / Copyright
Freedom of Information Act
Web Master: PDS-PPI Operator
NASA Official: Becky McCauley-Rench