VOLUME 1: MARS EXPRESS ASPERA3 EDR ELS DATA View Errata Order Volume Toggle viewing into brief mode
MEX-M-ASPERA3-2-EDR-ELS-EXT3-V1.0 PDS 3      CERTIFIED
Start Time: 2010-01-02 07:50:15 - Stop Time: 2013-01-01 13:29:19
Extension 3 raw data from the Electron Spectrometer (ELS) instrument of the Analyzer of Space Plasmas and Energetic Atoms experiment, 3rd edition (ASPERA-3), aboard the Mars Express spacecraft. The ASPERA-3 Principal Investigator is Dr. Rickard Lundin, Swedish Institute of Space Physics (IRF), Kiruna, Sweden. The primary source for ASPERA-3 data     more ...
Extension 3 raw data from the Electron Spectrometer (ELS) instrument of the Analyzer of Space Plasmas and Energetic Atoms experiment, 3rd edition (ASPERA-3), aboard the Mars Express spacecraft. The ASPERA-3 Principal Investigator is Dr. Rickard Lundin, Swedish Institute of Space Physics (IRF), Kiruna, Sweden. The primary source for ASPERA-3 data is the ESA Planetary Science Archive (PSA).    less ...
Citiation Information
Lundin, R., S. Barabash, D. Winningham, R. Frahm, MARS EXPRESS ASPERA-3 RAW EDR ELECTRON SPECTROMTR EXT3 V1.0, MEX-M-ASPERA3-2-EDR-ELS-EXT3-V1.0, ESA Planetary Science Archive (PSA), NASA Planetary Data System (PDS), 2005
Top/ DATA/ 2011/ 032_059FEB / click here to download folders and files
ELSSCIH20110522129C_ACCS01
Start Time: 2011-052T21:29:35.298 Stop Time: 2011-053T00:36:29.600
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20110530405C_ACCS01
Start Time: 2011-053T04:05:00.318 Stop Time: 2011-053T07:24:47.254
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20110531104C_ACCS01
Start Time: 2011-053T11:04:31.289 Stop Time: 2011-053T14:24:17.539
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20110531804C_ACCS01
Start Time: 2011-053T18:04:01.566 Stop Time: 2011-053T19:09:39.007
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20110531913C_ACCS01
Start Time: 2011-053T19:13:51.446 Stop Time: 2011-053T21:23:58.451
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20110540128C_ACCS01
Start Time: 2011-054T01:28:23.421 Stop Time: 2011-054T04:23:30.378
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20110540803C_ACCS01
Start Time: 2011-054T08:03:14.472 Stop Time: 2011-054T11:23:00.817
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20110541503C_ACCS01
Start Time: 2011-054T15:03:16.973 Stop Time: 2011-054T16:08:32.239
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20110541612C_ACCS01
Start Time: 2011-054T16:12:45.459 Stop Time: 2011-054T18:22:52.121
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20110542227C_ACCS01
Start Time: 2011-054T22:27:15.282 Stop Time: 2011-055T01:22:23.207
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20110550502C_ACCS01
Start Time: 2011-055T05:02:39.340 Stop Time: 2011-055T08:22:25.304
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20110551202C_ACCS01
Start Time: 2011-055T12:02:09.372 Stop Time: 2011-055T13:07:38.106
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20110551311C_ACCS01
Start Time: 2011-055T13:11:50.327 Stop Time: 2011-055T15:21:58.238
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20110551926C_ACCS01
Start Time: 2011-055T19:26:29.290 Stop Time: 2011-055T22:21:36.249
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20110560201C_ACCS01
Start Time: 2011-056T02:01:20.314 Stop Time: 2011-056T05:21:07.061
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20110560901C_ACCS01
Start Time: 2011-056T09:01:23.156 Stop Time: 2011-056T12:21:09.605
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20110561600C_ACCS01
Start Time: 2011-056T16:00:53.630 Stop Time: 2011-056T17:06:16.738
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20110561710C_ACCS01
Start Time: 2011-056T17:10:28.895 Stop Time: 2011-056T19:20:35.428
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20110562325C_ACCS01
Start Time: 2011-056T23:25:02.417 Stop Time: 2011-057T02:20:09.530
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20110570600C_ACCS01
Start Time: 2011-057T06:00:25.658 Stop Time: 2011-057T09:20:11.961
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20110571259C_ACCS01
Start Time: 2011-057T12:59:56.022 Stop Time: 2011-057T16:19:42.136
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20110571959C_ACCS01
Start Time: 2011-057T19:59:26.227 Stop Time: 2011-057T20:55:09.492
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20110572059C_ACCS01
Start Time: 2011-057T20:59:21.963 Stop Time: 2011-057T23:19:39.722
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20110580323C_ACCS01
Start Time: 2011-058T03:23:55.418 Stop Time: 2011-058T06:19:02.660
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20110580959C_ACCS01
Start Time: 2011-058T09:59:18.791 Stop Time: 2011-058T13:19:05.906
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20110581658C_ACCS01
Start Time: 2011-058T16:58:49.971 Stop Time: 2011-058T20:18:35.934
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20110582358C_ACCS01
Start Time: 2011-058T23:58:19.936 Stop Time: 2011-059T03:18:06.396
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20110590658C_ACCS01
Start Time: 2011-059T06:58:22.553 Stop Time: 2011-059T08:03:35.104
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20110590807C_ACCS01
Start Time: 2011-059T08:07:47.167 Stop Time: 2011-059T10:17:53.202
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20110591422C_ACCS01
Start Time: 2011-059T14:22:27.321 Stop Time: 2011-059T17:17:34.312
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIH20110592057C_ACCS01
Start Time: 2011-059T20:57:50.441 Stop Time: 2011-060T00:17:36.793
This file contains data from ELSSCIH:  Electron Spectrometer (ELS)
science data (SCI) from the high range (H) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20110522129C_ACCS01
Start Time: 2011-052T21:29:37.173 Stop Time: 2011-053T00:36:31.694
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20110530405C_ACCS01
Start Time: 2011-053T04:05:02.193 Stop Time: 2011-053T07:24:49.347
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20110531104C_ACCS01
Start Time: 2011-053T11:04:33.164 Stop Time: 2011-053T14:24:19.633
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20110531804C_ACCS01
Start Time: 2011-053T18:04:03.441 Stop Time: 2011-053T19:09:41.101
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20110531913C_ACCS01
Start Time: 2011-053T19:13:21.071 Stop Time: 2011-053T21:24:00.545
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20110540128C_ACCS01
Start Time: 2011-054T01:28:25.296 Stop Time: 2011-054T04:23:32.472
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20110540803C_ACCS01
Start Time: 2011-054T08:03:16.347 Stop Time: 2011-054T11:23:02.911
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20110541503C_ACCS01
Start Time: 2011-054T15:03:18.848 Stop Time: 2011-054T16:08:34.332
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20110541612C_ACCS01
Start Time: 2011-054T16:12:15.084 Stop Time: 2011-054T18:22:54.215
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20110542227C_ACCS01
Start Time: 2011-054T22:27:17.157 Stop Time: 2011-055T01:22:25.301
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20110550502C_ACCS01
Start Time: 2011-055T05:02:41.215 Stop Time: 2011-055T08:22:27.398
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20110551202C_ACCS01
Start Time: 2011-055T12:02:11.247 Stop Time: 2011-055T13:07:40.199
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20110551311C_ACCS01
Start Time: 2011-055T13:11:19.951 Stop Time: 2011-055T15:22:00.331
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20110551926C_ACCS01
Start Time: 2011-055T19:26:31.165 Stop Time: 2011-055T22:21:38.343
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20110560201C_ACCS01
Start Time: 2011-056T02:01:22.189 Stop Time: 2011-056T05:21:09.155
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20110560901C_ACCS01
Start Time: 2011-056T09:01:25.031 Stop Time: 2011-056T12:21:11.699
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20110561600C_ACCS01
Start Time: 2011-056T16:00:55.505 Stop Time: 2011-056T17:06:18.832
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20110561709C_ACCS01
Start Time: 2011-056T17:09:58.551 Stop Time: 2011-056T19:20:37.521
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20110562325C_ACCS01
Start Time: 2011-056T23:25:04.292 Stop Time: 2011-057T02:20:11.624
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20110570600C_ACCS01
Start Time: 2011-057T06:00:27.533 Stop Time: 2011-057T09:20:14.055
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20110571259C_ACCS01
Start Time: 2011-057T12:59:57.897 Stop Time: 2011-057T16:19:44.229
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20110571959C_ACCS01
Start Time: 2011-057T19:59:28.102 Stop Time: 2011-057T20:55:11.586
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20110572058C_ACCS01
Start Time: 2011-057T20:58:51.587 Stop Time: 2011-057T23:19:41.816
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20110580323C_ACCS01
Start Time: 2011-058T03:23:57.293 Stop Time: 2011-058T06:19:04.753
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20110580959C_ACCS01
Start Time: 2011-058T09:59:20.666 Stop Time: 2011-058T13:19:08.000
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20110581658C_ACCS01
Start Time: 2011-058T16:58:51.846 Stop Time: 2011-058T20:18:38.028
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20110582358C_ACCS01
Start Time: 2011-058T23:58:21.811 Stop Time: 2011-059T03:18:08.490
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20110590658C_ACCS01
Start Time: 2011-059T06:58:24.428 Stop Time: 2011-059T08:03:37.198
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20110590807C_ACCS01
Start Time: 2011-059T08:07:16.792 Stop Time: 2011-059T10:17:55.296
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20110591422C_ACCS01
Start Time: 2011-059T14:22:29.196 Stop Time: 2011-059T17:17:36.405
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ELSSCIL20110592057C_ACCS01
Start Time: 2011-059T20:57:52.316 Stop Time: 2011-060T00:17:38.887
This file contains data from ELSSCIL:  Electron Spectrometer (ELS)
science data (SCI) from the low range (L) of the ELS power supply
with up to 128 steps returned in telemetry.  These step values are
in the SCAN rows of the data files in the form of deflection
potential (volts).  To convert each step per sector to center
electron energies (eV), use the K_FACTORS in the calibration
tables (K-factor * defl. volt = eV). Each SCAN row directly
corresponds to the previous 16 SENSOR rows where the SENSOR value
(c/acc) in a given column is the data sampled at the SCAN value
(deflection voltage) in that same column.  The SENSOR rows have
one more column than the SCAN rows.  The additional column is the
data quality indicator (last column) for that sensor data.  The
data values are formatted for a high level of accuracy and DO NOT
reflect the accuracy of the intrumentation.  The ELS Instrument is
mounted on a scanning platform that was activated for the first
time in January 2006.  The scanner may have moved from its
parked position or be in motion for some of this data.  Please
refer to the Data Set Catalog, ASPERA3_EDR_ELS_DS.CAT, for
detailed information about the data organization, science meaning,
data quality determination, data modes, etc.
ICON LEGEND
For links to data holdings various icons are used to represent services or actions you can take. Those icons are:
Order the volume (CD-ROM/DVD) containing the listed data. The volume will be mailed to you. The volume (CD-ROM/DVD) contains some errors which are documented in an errata log.
Launch a custom web browser for the resource. Launch a map based browser for the resource.
Toggle viewing into brief mode (shorter descriptions are displayed). Toggle viewing into expanded mode (longer descriptions are displayed).
Download the volume containing the listed data.


NEW RELEASES

More.....

PDS RESOURCES

Archive Planning Guide
      Individuals
      Missions
Data Dictionary Search
      Lookup Tool
      Document
PDS4 Standards Documents
PDS Subscription Service

PDS3 Standards

OTHER RESOURCES

ADS Search
Science Data Licenses

Privacy / Copyright
Freedom of Information Act
Web Master: PDS-PPI Operator
NASA Official: Becky McCauley-Rench