LRO CRaTER Data Bundle                                                         
LRO Calibrated (Level 1) Data Collections (Housekeeping, Primary and Secondary)
                                                                               
Original PDS3_DATA_SET_ID           = LRO-L-CRAT-3-CDR-CALIBRATED-V1.0         
                                                                               
Original PDS3 DATA_SET_RELEASE_DATE = 2019-12-13                               
Version 2.4 Release Date            = 2026-09-15                               
START_TIME                          = 2009-06-29T00:00:00.00                   
STOP_TIME                           = 2026-06-30T23:59:59.01                   
PRODUCER_FULL_NAME                  = PROF. HARLAN SPENCE                      
                                                                               
Overview of Collections                                                        
=======================                                                        
                                                                               
    The Cosmic Ray Telescope for the Effects of Radiation (CRaTER) is a        
    stacked detector-absorber cosmic-ray telescope designed to answer key      
    questions to enable future human exploration of the Solar System.          
    CRaTER's primary measurement goal is to measure directly the average       
    lineal energy transfer (LET or 'y') spectra caused by space radiation      
    penetrating and interacting with shielding material.  Such measured LET    
    spectra are frequently unavailable.  In the absence of measurements,       
    numerical models are used to provide estimates of LET; the reliability of  
    the models require experimental measurements to provide a ground truth.    
                                                                               
    The calibrated (Level 1) data consists of files containing data processed  
    from the raw primary science, secondary science, and housekeeping data re- 
    cords.  During processing, the raw data are converted with instrument-     
    specific calibration and conversion factors to calibrated data records     
    containing science and engineering measurements and instrument oper-       
    ating parameters.  The calibrated data records are written to files in     
    plain text, fixed record format; each file contains calibrated data records
    for a single UTC day. All time values in Level 1 data products are in      
    spacecraft clock units.                                                    
                                                                               
    The Level 1 data are an intermediate data product meant to be used for     
    data processing diagnostics and troubleshooting.  Although the Level 1     
    collections can be used for some data analyses, it is not intended as the  
    primary source for further data analyses or scientific research. In the    
    Level 1 collections all times are expressed in spacecraft clock units;     
    spacecraft location and instrument pointing data are not included. Users   
    seeking CRaTER data are instead encouraged to use the Level 2 derived data 
    record collections.  The Level 2 data contain all Level 1 data supplemented
    time values converted to UTC and computed spacecraft location and          
    instrument pointing information.                                           
                                                                               
    See [Spence et al. 2010] for detailed description of LRO flight version    
    of the instrument, its operations, and data processing.                    
                                                                               
    Science Objectives and Observation Strategy                                
    -------------------------------------------                                
    CRaTER is designed to achieve characterization of the global lunar         
    radiation environment and its biological impacts and potential mitigation  
    as well as investigation of shielding capabilities and validation of       
    other deep space radiation mitigation strategies involving materials.      
    CRaTER will fill knowledge gaps regarding radiation effects, provide       
    fundamental progress in knowledge of the Moon's radiation environment,     
    and provide specific path-finding benefits for future planned human        
    exploration.                                                               
                                                                               
    Parameters                                                                 
    ----------                                                                 
    LRO CRaTER flight instrument identification:                               
    --instrument model = Flight Model 1 (FM1);                                 
    --instrument serial number (S/N) = 02;                                     
    --FPGA revision code = 3.                                                  
                                                                               
    Data                                                                       
    ----                                                                       
    CRaTER's principal measurement is the energy deposited in the 3 pairs of   
    silicon detectors by charged particles and photons passing through the     
    instrument's 'telescope' unit.  Whenever the coulombic charge signal re-   
    sulting from the energy deposited in a detector exceeds a predefined and   
    fixed threshold, the instrument's electronics performs a detailed          
    measurement of the signals from all of the detectors.  The resulting       
    detector signal amplitudes are compared to the values of the 'lower level  
    discriminators' (LLDs).  LLDs establish minimum amplitudes for signals to  
    qualify as valid charged-particle or photon interactions.  The LLD values  
    are generally set to insure that the desired charged-particle or photon    
    measurements are not contaminated by system electronic noise.  Separate    
    LLD settings are required for the thick and thin detectors due to the      
    difference in their sensitivities; the thin and thick detector LLD values  
    are reported in the 'DiscThin' and 'DiscThick' parameters as part of the   
    secondary science packet.                                                  
                                                                               
    In addition to the LLD settings, measurement filtering is achieved         
    through detector coincidence requirements--the combination of detectors    
    registering valid signals to qualify as a charged-particle or photon       
    measurement 'event'.  To measure all charged particles arriving from the   
    instrument's zenith or nadir directions, for example, the coincidence      
    requirements would be valid signals in at least detectors 1, or 2, or 5,   
    or 6.  Conversely, a coincidence consisting of valid signals in all six    
    detectors would ensure only zenith- or nadir-arriving charged particles    
    with high energies are reported.  For CRaTER's six axially-coaligned       
    detectors there are 64 possible coincidence combinations.  The desired     
    set of coincidence combinations are stored as a coincidence mask           
    parameter in the instrument's memory; the coincidence mask setting is      
    reported in the 'Mask' parameter as part of the secondary science packet.  
                                                                               
    To qualify as an 'event', therefore, a charged particle or photon passing  
    through CRaTER's telescope must interact and deposit sufficient energy to  
    generate signals with amplitudes in excess of the specified LLDs in a      
    specified combination of detectors; only data for valid 'events' are re-   
    ported in the instrument's telemetry.                                      
                                                                               
    The measured interaction event data is written as a series of primary      
    science packets to the instrument's output telemetry buffer for the        
    spacecraft to read. At ~1 second intervals CRaTER receives a timing pulse  
    from the spacecraft, at which time it flushes the primary science data     
    from the output buffer and writes a secondary science packet for the       
    spacecraft to read.  Every 16 seconds a housekeeping packet is also        
    created and written to the output buffer.                                  
                                                                               
    The Level 1 data are created from the corresponding Level 0 data by        
    converting the instrument binary output with conversion and calibration    
    factors to science and engineering data.                                   
                                                                               
    The Level 1 collections are composed of the three types of time-sequential 
    calibrated data records: (1) primary science, (2) secondary                
    science, and 3) housekeeping.  The three types of calibrated data records  
    are written to separate data files in plain text, fixed record format.     
    Each file contains calibrated data records for a single UTC day.           
                                                                               
    The Level 1 primary science data consists of a sequence of interaction     
    event calibrated data records--one CDR for each measured event.  Each      
    calibrated data record consists of the energy deposited in each of the six 
    detectors and the spacecraft time at the end of the measurement interval   
   (receipt of spacecraft timing pulse). calibrated data records for events    
    recorded in the same measurement interval have the same time tags--the     
    'SECONDS' and 'FRACT' field values.  Although numerous events may have the 
    same time value, the events are recorded in the order in which they        
    occurred; this relative order is captured in the calibrated data records   
    'INDEX' field.                                                             
                                                                               
    The Level 1 secondary science calibrated data records contain the majority 
    of instrument configuration settings, status flags, and event counters.    
    Reported configuration settings include the last command sent to CRaTER,   
    detector LLD settings, and coincidence mask values.  Status flags available
    in the secondary science calibrated data records include detector bias     
    status, selected pulse amplitude range and rate for the internal           
    calibration pulser, and detector processing status.  Counters report the   
    number of 'singles' for each detector as well as the number of 'good',     
    'rejected', and total events recorded by CRaTER during the monitoring      
    period.                                                                    
                                                                               
    The Level 1 housekeeping calibrated data records contain measured          
    instrument operating and environmental parameters used to assess the health
    and performance of the instrument, such as power supply output voltages,   
    detector bias voltages and currents, pulse amplitudes from the internal    
    calibration pulser, and temperatures at five locations inside of the       
    instrument's housing. The analog output signal (voltage) from radiation    
    monitor is also included the housekeeping calibrated data records.         
                                                                               
    Confidence Level Overview                                                  
    -------------------------                                                  
    An assessment of the accuracy and precision of data in the LRO CRaTER      
    calibrated (Level 1) collections is limited to the measured deposited      
    energy in each detector.  General instrument housekeeping param-           
    eters (e.g., temperatures, voltages, currents, LLD voltages, pulser        
    signal amplitudes, spacecraft clock value) are provided with no statement  
    of uncertainty--the accuracy of these parameters is assumed to be suf-     
    ficient for general correlation and trending analysis.  The accuracy of    
    the housekeeping temperature parameters has an impact on the accuracy and  
    precision of the conversion from detector PHA channel numbers to de-       
    posited energy values; this impact, however, is very small in comparison   
    to other sources of systematic and stochastic error.                       
                                                                               
    Potential sources of instrument systematic error include signal pulse      
    shaping output linearity, analog-to-digital conversion (ADC) linearity,    
    electronic calibration source stability and linearity, and the accuracy    
    of the gain and offset values determined for each detector-amplifier-ADC   
    string.                                                                    
                                                                               
    The linearity of the amplifier-ADC strings (i.e., pulse height             
    analyzer or PHA) was established with a precision external pulser.  For a  
    given pulser output setting, the variability in output pulse amplitude is  
    determined to be 0.01%.  Over the pulser's full range of output pulse      
    amplitude settings, the measured pulse amplitudes were found to be very    
    linear, with an RMS fit residual upper limit of 0.1%.                      
                                                                               
    The external pulser was used to establish the linearity of the six CRaTER  
    PHA circuits.  The precision external pulser served as a calibrated input  
    charge source by coupling it (via a precision capacitor) to the base of    
    each PHA circuit's preamplifier.  Each PHA circuit's response was found    
    to be very linear, with RMS fit residuals significantly less than 0.1%.    
                                                                               
    Temporal stability of the PHA circuits was established through repeated    
    testing with the external pulser over an 15-month period.  Between Sep     
    2007 and Jan 2009, each PHA circuit was tested five times at a fixed       
    pulser output setting.  The output of each PHA circuit was determined to   
    be very stable, with ~0.06% variability in the value of the center of the  
    PHA peak.                                                                  
                                                                               
    Temperature dependence of the gain of each PHA circuit was measured over   
    the expected range of operating temperatures during the LRO mission.  The  
    output of each PHA circuit to fixed amplitude pulses from the precision    
    external pulser was measured with the CRaTER instrument operating at -30   
    degrees C, -10 degrees C, +10 degrees C, and +35 degrees C (temperature    
    measured inside the instrument's case close to the analog and digital      
    circuit boards).  The PHA circuit gains were found to be fairly stable     
    over this temperature range, with only a weak non-linear temperature de-   
    pendence.  Detectors 2, 4, and 6 PHA circuits exhibited gain variations    
    of ~ +/- 0.1% over the temperature range; detectors 1, 3, and 5 PHA        
    circuits gains varied by ~ +/- 0.5%.                                       
                                                                               
    Potential sources of stochastic error include electronic noise, uncer-     
    tainty in the PHA-channel-to-deposited-energy conversion factors (i.e.,    
    'calibration values'), and uncertainty in actual deposited energy values   
    due to digitization.                                                       
                                                                               
    From the standard deviations of the pulse amplitudes measured over the     
    full dynamic range of each amplifier-A-to-D-converter strings, the upper   
    limit on system electronic noise is approximately 0.15% of pulse           
    amplitude or 0.02% of each string's maximum output value. [The system      
    electronic noise measured with CRaTER operating at 10 degrees C].          
                                                                               
    PHA channel number is converted to deposited energy by                     
                                                                               
      Ei [keV] = GiCi + Oi, where                                              
                                                                               
      Ei [keV] =              deposited energy measured by detector/           
                              PHA chain i,                                     
      Ci [ADU or channel #] = output from detector/PHA chain i,                
      Gi [keV/ADU] =          gain of detector/PHA chain i, and                
      Oi [keV] =              offset of detector/PHA chain i.                  
                                                                               
    The calibration values Gi and Ci used to convert PHA output to deposited   
    energy were determined through a combination of alpha particle exposure    
    measurements and modeling of the instrument's response to moderate energy  
    protons.  A more extensive description of the calibration process is       
    found in [Spence et al. 2010].                                             
                                                                               
    The LRO CRaTER instrument V1.0 calibration values are listed in            
    [Spence et al. 2010], table 6, and reproduced here.                        
                                                                               
    Parameter      Units      D1     D2     D3     D4     D5     D6            
    -----------------------------------------------------------------          
    Gain, Gi       keV/ADU   76.3   21.8   78.6   21.6   76.3   21.9           
    Offset, Oi     keV       105.1  50.0   152.8  74.7   119.1  46.6           
                                                                               
   The uncertainty in the Gi and Ci values awaits further analysis.            
                                                                               
    The process of converting the detector signals into digital values re-     
    quires discretizing the amplifier analog output signals into one of a      
    possible 4096 linearly-spaced values.  These 4096 'channel' or 'ADU'       
    values correspond to ranges of ~0-300 MeV and ~0-90 MeV for the thin and   
    thick detector PHA circuits, respectively.  Each PHA channel corresponds   
    to a small but finite range of energies described by a probability         
    distribution rather than a discrete energy value.  The calibration         
    process establishes an effective energy and energy width for each          
    channel. Assuming the actual deposited energy probability distribution     
    for a given PHA channel is approximately flat, the average energy and      
    uncertainty corresponding to the channel are the effective energy and      
    energy width established through calibrations.  While the absolute         
    magnitude of the uncertainty resulting from discretization is a constant   
    value (one-half the gain), the relative uncertainty is a function of the   
    energy corresponding to the particular PHA channel--the lower the          
    channel's corresponding energy, the higher the realtive uncertainty.       
    The discretization uncertainty extremes are summarized in the              
    following table.                                                           
                                                                               
    Detector/         Energy (keV)              Energy (keV)                   
    PHA Chain          PHA = 0 ADU               PHA = 4095                    
    ----------------------------------------------------------                 
    D1           105.1 +/- 38.2 (36.3%)    312554 +/- 38.2 (0.012%)            
    D2           50.0  +/- 10.9 (21.8%)    89321  +/- 10.9 (0.012%)            
    D3           152.8 +/- 39.3 (25.7%)    322020 +/- 39.3 (0.012%)            
    D4           74.7  +/- 10.8 (14.5%)    88527  +/- 10.8 (0.012%)            
    D5           119.1 +/- 38.2 (32.0%)    312568 +/- 38.2 (0.012%)            
    D6           46.6  +/- 11.0 (23.5%)    89727  +/- 11.0 (0.012%)            
                                                                               
    For PHA values ❯ 48 ADU, the relative uncertainty in the deposited energy  
    due to discretization is ❮ 1% a for all detector/PHA chains.               
                                                                               
    This overview has identified, described, and where possible enumerated the 
    various error/uncertainty components.  The confidence levels for the total 
    cumulative uncertainty in the measured deposited energies values awaits    
    further analysis.  When the values become available a revision will be     
    provided to this catalog file.                                             
                                                                               
    Review                                                                     
    ------                                                                     
    A minimal set of automated quality control steps are used by the data      
    processing system to verify the integrity of the data during the initial   
    creation of the raw data files.  Each raw data packet's CCSDS header is    
    checked for format and content. Packets are discarded if their headers     
    are corrupted, incorrectly formatted, or containing invalid values.  All   
    packets are sorted into time order and checked for temporal gaps.  Dupli-  
    cate packets are also discarded.  Metrics plus any detected anomalies are  
    written to process log files for review by scientists and engineers from   
    the instrument team.  Anomalies noted during the processing are investi-   
    gated.  Anomalies due to missing input files (e.g., instrument science     
    and housekeeping data files, spacecraft housekeeping data files,           
    spacecraft ephemeris kernels, and ancillary files such as leap second and  
    spacecraft clock kernels) are corrected by locating the missing input and  
    reprocessing the data.                                                     
                                                                               
    All data is periodically analyzed using graphical and statistical methods  
    to check for out-of-range values as well as anomalous trends that may      
    indicate detector and/or amplifier-ADC string degradation.                 
                                                                               
    Data Coverage and Quality                                                  
    -------------------------                                                  
    The start date for the initial version of the LRO CRaTER Calibrated (Level 
    1) archival data is 2009-06-29T00:00:00.000.  This date/time is the        
    beginning of the first full day following completion of LRO lunar orbit    
    insertion (LOI) and transition to the nominal nadir-pointing observation   
    attitude.  It is also the first day for which complete re-constructed      
    ephemeris ('SPK') data was provided by the LRO Mission Operations Center.  
                                                                               
    Data gaps are identified during initial data processing. The gap start and 
    stop times are recorded in gap files stored in the document directory --   
    there are separate gap files for the primary science, secondary science,   
    and housekeeping data sets.  Each gap file contains a cumulative listing   
    of the missing data up to and including the days for the data current vol- 
    ume.  Description of overall data coverage and quality. This section       
    should include information about gaps in the data (both for times or re-   
    gions) and details regarding how missing or poor data are flagged or       
    filled, if applicable.  The minimum duration between successive data       
    packets to qualify as a data gap is specified during data processing. The  
    default durations are 2 seconds for both primary and secondary science     
    data packets, and 20 seconds for housekeeping data packets.  These values  
    may be over ridden at the time of data processing, however.  The actual    
    durations used while processing a specific set of data are recorded in the 
    corresponding process log file.                                            
                                                                               
    Aperiodic episodes of sporadic, significant elevation in the thick detec-  
    tor (D2, D4, and D6) singles rates have been observed during all phases    
    of mission phases. The elevated singles rates most commonly occur in       
    detector D2, but have also been observed in detector D6; a detector's      
    singles rate may increase by a factor of 20 or more.  During these         
    periods increases may occur in both the 'reject' and 'good' event rates.   
    Episodes tend to last for three to five weeks, followed by extended        
    periods with nominal singles rates.  During an episode singles rates       
    vary sporadically between nominal and extremely elevated levels, although  
    there seems to be a general gradual build-up and decline in the peak       
    magnitude of the singles rates over the course of an episode.  Despite     
    intensive analysis, the cause for the periods of elevated singles rates    
    has not yet been determined.  No correlation has been found with           
    spacecraft location, local space and spacecraft environment conditions,    
    instrument boresite direction, or spacecraft and instrument operations.    
    Users are urged to first plot the detector singles rates and 'good' and    
    'reject' event rates as a function of time to identify periods with        
    elevated singles rates which may impact their particular use of the data.  
                                                                               
    Limitations                                                                
    -----------                                                                
    The LRO CRaTER Calibrated (Level 1) data include all Level 1 data obtained 
    by the CRaTER instrument, including data from periods when the instrument  
    was placed into special configurations.  Special configurations include    
    the instrument start-up tests that occur whenever the instrument is power  
    cycled to (e.g., initial instrument start-up, recovery following space-    
    craft transition to sun-safe mode) as well routine calibrations (90-degree 
    off-nadir GCR background measuerments, internal pulser sweeps, LLD zero    
    crossing measurements, and LLD sweeps).  These periods can be detected by  
    monitoring the 'CalLow' and 'CalHigh' flags and 'DiscThin' and 'DiscThick' 
    LLD values in the secondary science calibrated data records.               
                                                                               
    Timing resolution for the set of events recorded between two successive    
    timing pulses (buffer readouts)is limited to the corresponding spacecraft  
    times.  If, for example, 560 particle 'events' are measured between two    
    successive timing pulses, the exact time of each event's occurrence is     
    unknown--all that is known is that event was measured between the times    
    of the two timing pulses. The sequence in which the events were measured,  
    however, is preserved-for a given time interval, the first reported event  
    was measured before the second reported event, etc.                        
                                                                               
    The maximum rate at which detector measurements can be reported in the     
    primary science data is ~1200 events per second; the true number of        
    events in each time interval is reported in the secondary science derived  
    data records.                                                              
                                                                               
    Users should be aware of the impact of the LLD settings on the primary     
    and secondary science data.  The LLD settings establish the minimum        
    amplitudes of the amplifier output pulse heights (i.e., minimum deposited  
    energies) to qualify as a valid signal and trigger the ADC process.  In    
    addition to determining the lower limit of the PHA and LET spectra, the    
    choice of LLD values directly affects the number of 'good' and 'reject'    
    events reported in the secondary science data derived data records. For a  
    given set of incident charged-particle energy spectra, as the LLD values   
    increase, the 'good' and 'reject' event rates will decrease. Users         
    analyzing the temporal variability of 'good' and 'reject' event rates      
    should ensure the LLD settings do not change over the analysis period.     
    The nominal instrument operating mode maintains constant LDD settings.     
    Modes using varying LLD settings, however, occur during instrument power-up
    tests and routine calibration procedures.  In addition, as the mission     
    progresses changes in noise levels due to instrument component aging may   
    require adjustments to the baseline LLD settings.                          
                                                                               
Reference                                                                      
=========                                                                      
                                                                               
    [Spence et al. 2010] Spence, H.E., A.W. Case, M.J. Golightly,              
    T. Heine, B.A. Larsen, J.B. Blake, P. Caranza, W.R. Crain, J. George,      
    M. Lalic, A. Lin, M.D. Looper, J.E. Mazur, D. Salvaggio, J.C. Kasper,      
    T.J. Stubbs, M. Doucette, P.Ford, R. Foster, R. Goeke, D. Gordon,          
    B. Klatt, J. O'Conner, M. Smith, T. Onsager, C. Zeitlin, L.W. Townsend,    
    Y. Charara (2010), CRaTER:  The Cosmic Ray Telescope for the Effects of    
    Radiation Experiment on the Lunar Reconnaissance Orbiter Mission, Space    
    Sci. Rev., 150, 243-284, DOI: 10.1007/s11214-009-9584-8.